SensoLite Nova
Checking with Check-strip
Press OK button and keep it pressed until “CAL” appears on the
screen, then release it.
Strip icon will be displayed together with flashing arrow icon,
indicating that the meter is waiting for strip insertion.
Insert Check-strip – which is included in the package of
SensoLite Nova – into strip holder with the “Check” label
facing upwards. Meter will perform the measurement and
display the result of the check.